10 - On
Time, Adjustable, 2~90 seconds, 1 Sec. / step.
11 - Off
Time, Adjustable, 0~90 seconds, 1 Sec. / step.
12 - Ramp
Time, Adjustable, 1~8 seconds, 1 Sec. / step,
The "On" time will increase and decrease in the setting value.
13 - Modes,
Five TENS Modes: B (Burst), N (Normal), M (Modulation), SD1 &
SD2 (Strength Duration),
14 - Burst
Mode (B), Burst rate: Adjustable, 0.5 ~ 5Hz Pulse width
adjustable, 50~300μs. Frequency fixed = 100 Hz.
15 - Normal
Mode (N), Pulse rate and pulse width are adjustable.
It generates continuous stimulation based on the setting value.
16 - Modulation
Mode (M). Modulation mode is a combination of pulse rate and
pulse width modulation. The pulse rate and width are
automatically varied in a cycle pattern. The pulse width is
decreased by 50% from its original setting in 0.5 second, and
then the pulse rate is decreased by 50% from its original
setting in 0.5 second. Total cycle time is 1 second. In this
mode, pulse rate (2-150Hz) and pulse width (50-300μs) are fully
adjustable.
17 - SD1
Mode, The SD1 (Strength-Duration) mode consists of automatic
modulation intensity and pulse width in 40% range.
The intensity is always increasing while the pulse width is
decreasing and vice-versa. The intensity is decreased by 40%
while the pulse width is increased by 40% in 5 seconds. In the
next 5 seconds, the intensity is increased by 40% while the
pulse width is decreased by 40%.Total cycle time is 10 seconds.
Pulse rate (2~150Hz) and pulse width (50~300μs) are fully
adjustable.
18 - SD2
Mode, The SD2 (Strength-Duration) mode consists of automatic
modulation intensity and pulse width in 70% range. The
intensity is always increasing while the pulse width is
decreasing and vice-versa. The intensity is decreased by 70%
while the pulse width is increased by 70% in 5 seconds. In the
next 5 seconds, the intensity is increased by 70% while the
pulse width is decreased by 70%.Total cycle time is 10 seconds.
Pulse rate (2~150Hz) and pulse width (50~300μs) are fully
adjustable.
19 - Constant
Mode (C), Constant stimulation based on setting value.
Only pulse width,
pulse rate and timer are adjustable in this mode. "Constant" is
equal o the “Normal" mode of a TENS unit.
20 - Synchronous
Mode (S), Stimulation of both channels occurs synchronously. The
"ON" time including "Ramp Up" and “Ramp Down" time. Therefore,
the setting of ON Time should be no less than two times of the
"Ramp" time in this mode. ON TIME
≧Ramp
up + Ramp down.
21 - Alternate
Mode (A), the stimulation of the CH2 will occur after the 1st
contraction of CH1 is completed. In this mode, the setting of ON
Time should be no less than two times of the "Ramp" time. The
OFF Time should be equal or more than the ON Time. ON TIME >Ramp
up + Ramp down OFF TIME
≧ON
TIME
22 - Timer
Two Adjustable timers, from 1 to 60 minutes or Continuous.
Adjustable in 1 minute each step from 1 to 15 minutes, and 5
minutes each step from 15 to 60 minutes. Treatment time
countdown automatically.
23 - Patient
Compliance Meter, This unit can store 60 sets of operation
records. Total recorded time is 999 hours.
24 - Low
Battery Indicator, A low battery indicator will show up when the
battery is low.
25 - Operating
Condition, Temperature: 0℃~40℃Relative
Humidity: 30%~75%