2 - Pulse
Amplitude, Adjustable, 0-100 mA peak into 500 ohm load each
channel.
3 - Wave
Form, Asymmetrical Bi-Phase
Square Pulse.
4 - Voltage,
0 to 50V (Load: 500 ohm).
5 - Power
Source, One 9V battery and / or mains power adaptor.
6 - Size,
11.8cm (L) x 6.1cm (W) x 3.1cm (H).
7 - Weight, 157
grams with battery.
8 - Pulse
Rate, Adjustable, from 2 to 150 Hz, 1 Hz/step.
9 - Pulse
Width, Adjustable, from 50 to 300 microseconds, 10 μs / step.
10 - On
Time, Adjustable, 2~90 seconds, 1 Sec. / step.
11 - Off
Time, Adjustable, 2~90 seconds, 1 Sec. / step.
12 - Ramp
Time, Adjustable, 1~8 seconds, 1 Sec. / step,
The "On"
time will increase and decrease in the setting value.
13 - Modes,
Five TENS Modes: B (Burst), N (Normal),
M (Modulation), S1 & S2 (Strength Duration),
14 - Burst
Mode (B), Burst rate: Adjustable, 0.5 ~ 5Hz Pulse width
adjustable, 50~300μs. Frequency fixed = 100 Hz.
15 - Normal
Mode (N), Pulse rate and pulse width are adjustable. It generates
continuous stimulation based on the setting value.
16 - Modulation
Mode (M). Modulation mode is a combination of pulse rate and
pulse width modulation. The pulse
rate and width are automatically varied in a cycle pattern. The
pulse width is decreased by 50% from its original setting in 0.5
second, and then the pulse rate is decreased by 50% from its
original setting in 0.5 second. Total cycle time is 1 second. In
this mode, pulse rate (2-150Hz) and pulse width (50-300μs) are
fully adjustable.
17 - S1
Mode, The S1 (Strength-Duration) mode consists of automatic
modulation intensity and pulse width in 40% range.
The
intensity is always increasing while the pulse width is
decreasing and vice-versa. The intensity is decreased by 40%
while the pulse width is increased by 40% in 5 seconds. In the
next 5 seconds, the intensity is increased by 40% while the
pulse width is decreased by 40%.Total cycle time is 10 seconds.
Pulse rate (2~150Hz) and pulse width (50~300μs) are fully
adjustable.
18 - S2
Mode, The S2 (Strength-Duration) mode consists of automatic
modulation intensity and pulse width in 70% range.
The
intensity is always increasing while the pulse width is
decreasing and vice-versa. The intensity is decreased by 70%
while the pulse width is increased by 70% in 5 seconds. In the
next 5 seconds, the intensity is increased by 70% while the
pulse width is decreased by 70%.Total cycle time is 10 seconds.
Pulse rate (2~150Hz) and pulse width (50~300μs) are fully
adjustable.
19 - Constant
Mode (C), Constant stimulation based on setting value.
Only pulse
width, pulse rate and timer are adjustable in this mode.
"Constant" is equal to the “Normal"
mode of a TENS unit.
20 - Synchronous
Mode (S), Stimulation of both channels occurs synchronously. The "ON"
time including "Ramp Up" and “Ramp Down" time. Therefore, the
setting of ON Time should be no less than two times of the
"Ramp" time in this mode. ON TIME ≧Ramp
up + Ramp down.
21 - Alternate
Mode (A), the stimulation of the CH2 will occur after the 1st
contraction of CH1 is completed. In this
mode, the setting of ON Time should be no less than two times of
the "Ramp" time. The OFF Time should be equal or more than the
ON Time. ON TIME >Ramp up + Ramp down OFF TIME ≧ON
TIME
22 - Timer
Adjustable timer, from 5 to 60 minutes or Continuous. Adjustable
in 5 minutes each steps.
23 - Patient
Compliance Meter, This unit can store 60 sets of operation
records. Total
recorded time is 999 hours.
24 - Low
Battery Indicator, A low battery indicator will show up when the
battery is low.